Saga device release notes v2.5.x
Solution/product | Saga device |
Version | 2.5.0 |
Change control reference | Page ID: 1418494016 |
Reason for change |
|
Type of release | Minor |
Release date | 31 January 2024 |
Customer actions needed | To get the firmware update, make sure that the devices are charged, not in a shipment, and have a connection to the cloud. |
Validation | This release will require partial validation of the device. |
New feature
The battery cycling feature will help maintain good battery health when devices are serviced in the Controlant Service Centers and will identify devices with potentially sub-optimal battery capacity.
Improvements
Improved battery life in low signal conditions by separating syncing to the cloud from registering onto a cellular network.
Improved remaining battery capacity estimation.
Local DNS caching in case of global DNS outage.
Increased maximum button push time from 12 seconds to 20 seconds.
More robust AD offset management in the firmware.
Other general stability improvements.
Bug fixes
Fixed a bug that caused some devices to lose their configuration.
Fixed a bug that caused some devices to resend old data after rebooting.
Fixed a bug that sometimes caused unexpected reboots due to stack overflow.