Maintenance

In this section:

To ensure the smooth operation of Clinical Storage Monitoring, take care to follow these steps:

Replace devices every 10 months

Saga devices need to be calibrated annually, so for each device, every 10 months, Controlant will do the following:

  1. Send you a replacement device.

  2. Send you instructions for returning the existing device for calibration.

Switch out devices

When you replace a Saga device for a storage unit, you need to do the following:

  1. When you have received a replacement device, do the following:

    1. Create a new shipment in Supply Chain Monitoring platform.

    2. Link the new device to the shipment.

    3. Start the shipment.

  2. When you have confirmed the new shipment is started, do the following:

    1. On the device you are replacing, click Stop.

      The shipment stops.

    2. Return the device to Controlant.